Question
I tried to measure the Voltage Noise of my power supply unit which is specified to the following values:
I tried to measure the Voltage Noise of my power supply unit which is specified to the following values:
My results have been much higher and I sent in the device for alignment/calibration. After getting it back I still gather the same results. Can you explain how to perform such a measurement?
We set up an example measurement with the following setup (ref. to IEEE Std 1515-2000):
Measurement 1: PSU Mains OFF
Measurement 2: PSU Mains ON, Output CH1 OFF:
Measurement 3: PSU Mains ON, CH1 Output ON:
The RMS value for CH1 is in specs.
Remarks for voltage mode:
Following the standard, the specified value (datasheet) has been captured at 90% load (current). In this case, the basis is a specific application running with about 1 A at a 10 Ohm load. We recommend performing the measurement only with a total ohmic load as electronical loads can cause additional noise!